Preview

Russian Sklifosovsky Journal "Emergency Medical Care"

Advanced search

Germicidal Efficiency of 272 nm LED in Relation to the Hospital Strain of Klebsiella Pneumoniae

https://doi.org/10.23934/2223-9022-2023-12-3-376-385

Abstract

Relevance. Currently, the use of ultraviolet (UV) radiation for the disinfection of objects and the treatment for infectious diseases is considered as a promising alternative to chemical biocides and antibiotics. Shortwave — UV-C and UV-B -light emitting diodes (LED) are a relatively new type of UV radiation sources and potentially able to meet the requirements of current medical technologies. However, their possibilities for the treatment of wounds and infectious diseases have not been practically researched to date, which determines the relevance of experiments aimed at studying the biocidal and therapeutic properties of shortwavelength UV LEDs.

Purpose. OF STUDY To evaluate the bactericidal efficacy of 272 nm LED radiation against hospital strains of Klebsiella pneumoniae bacteria characterized by multidrug resistance.

Materials and Methods. The studies were carried out with an experimental sample of the LED apparatus for UV irradiation. In the irradiator of the device, 5 LEDs are installed with a maximum radiation at a wavelength of 272 nm and a total electrical power of 10 watts. The UV radiation dose achieved in one irradiation session (12 seconds) at a distance of 10 cm from the irradiator was 8 mJ/cm2 . In the experiments, a hospital strain of the bacterium Klebsiella pneumoniae, isolated from the patient’s blood, was used. The strain was characterized by multidrug resistance. A daily culture suspension of K. pneumoniae with a concentration of 108 CFU/ml in a volume of 100 µl was transferred into a Petri dish with a diameter of 9 cm with meat-peptone agar and evenly distributed over a surface with a diameter of 8 cm. Petri dishes were irradiated from a distance of 10 cm from the irradiator. The change in the dose of UV irradiation from 4 to 80 mJ/cm2 was carried out by varying the exposure time. Studies were carried out in 4 repetitions at each dose. After irradiation, the experimental and control (without irradiation) Petri dishes were placed in a thermostat at 37ºC for 24 hours, then the grown colonies were counted. A total of 60 experiments were carried out.

Results. As a result of the research, it was shown that the LED device based on five 272 nm diodes provides deep and prompt disinfection of the surface from hospital strains of K. pneumoniae bacteria characterized by multidrug resistance. A dose of UV radiation of 8 mJ/cm2 reduces surface contamination with K. pneumoniae bacteria by more than a million times (decontamination efficiency over 99.9999%). At doses less than 10 mJ/cm2 , the efficiency of the 272 nm LED device in terms of inactivation of K. pneumoniae bacteria is 3–4 times higher than the bactericidal efficiency of mercury lamps.

Conclusion.The prospects of using UV devices based on LEDs with a maximum radiation at a wavelength of 272 nm in systems for the operational disinfection of massively contaminated surfaces, potentially including wound surfaces, have been shown.

About the Authors

A. S. Kamrukov
Bauman Moscow State Technical University
Russian Federation

Alexander S. Kamrukov - Candidate of Technical Sciences, Associate Professor, Head of Department, Research Institute of Power Engineering, 

bldg. 1, 5, 2nd Baumanskaya Str., 105005, Moscow



T. V. Chernenkaya
N.V. Sklifosovsky Research Institute for Emergency Medicine
Russian Federation

Tatyana V. Chernenkaya Candidate of Medical Sciences, Head, Scientific Laboratory of Clinical Microbiology, 

3, B. Sukharevskaya Sq., 129090, Moscow



L. Yu. Volodin
Bauman Moscow State Technical University
Russian Federation

Lev Yu. Volodin - Leading Engineer, Research Institute of Power Engineering,

bldg. 1, 5, 2nd Baumanskaya Str., 105005, Moscow



S. S. Petrikov
N.V. Sklifosovsky Research Institute for Emergency Medicine
Russian Federation

Sergey S. Petrikov - Corresponding Member of the Russian Academy of Sciences, Doctor of Medical Sciences, Professor, Director, 

3, B. Sukharevskaya Sq., 129090, Moscow



K. A. Popugaev
N.V. Sklifosovsky Research Institute for Emergency Medicine
Russian Federation

Konstantin A. Popugaev - Doctor of Medical Sciences, Professor, Deputy Director, 

3, B. Sukharevskaya Sq., 129090, Moscow



V. V. Bagrov
Bauman Moscow State Technical University
Russian Federation

Valery V. Bagrov - Candidate of Technical Sciences, Deputy Director, Research Institute of Power Engineering, 

bldg. 1, 5, 2nd Baumanskaya Str., 105005, Moscow



I. V. Bukhtiyarov
Izmerov Research Institute of Occupational Health
Russian Federation

Igor V. Bukhtiyarov - Corresponding Member of the Russian Academy of Sciences, Doctor of Medical Sciences, Professor, Director, 

31 Budyonny Ave., 105275, Moscow



E. V. Zibarev
Izmerov Research Institute of Occupational Health
Russian Federation

Evgeny V. Zibarev - Candidate of Medical Sciences, Deputy Director, 

31 Budyonny Ave., 105275, Moscow



K. A. Semenov
Bauman Moscow State Technical University bldg. 1, 5, 2nd Baumanskaya Str., 105005, Moscow
Russian Federation


V. I. Krylov
Bauman Moscow State Technical University bldg. 1, 5, 2nd Baumanskaya Str., 105005, Moscow
Russian Federation


References

1. Gupta A, Avci P, Dai T, Huang YY, Hamblin MR. Ultraviolet Radiation in Wound Care: Sterilization and Stimulation. Adv Wound Care. 2013;2(8):422–437. https://doi.org/10.1089/wound.2012.0366 PMID: 24527357

2. Muramoto Y, Kimura M, Nouda S. Development and future of ultraviolet light emitting diodes: UV-LED will replace the UV lamp. Semicon Sci Technol. 2014;29(8):084004. https://doi.org/10.1088/02681242/29/8/084004

3. Rattanakul S, Oguma K. Inactivation kinetics and efficiencies of UVLEDs against Pseudomonas aeruginosa, Legionella pneumophila and surrogate microorganisms. Water Res. 2018;130:31–37. https://doi.org/10.1016/j.watres.2017.11.047 PMID: 29195159

4. Nishisaka-Nonaka R, Mawatari K, Yamamoto T, Kojima M, Shimohata T, Uebanso T, et al. Irradiation by ultraviolet light- emitting diodes inactivates influenza a viruses by inhibiting replication and transcription of viral RNA in host cells. J Photochem Photobiol B. 2018;189:193–200. https://doi.org/10.1016/j.jphotobiol.2018.10.017 PMID: 30391908

5. Cheng Y, Chen H, Sánchez Basurto LA, Protasenko VV, Bharadwaj S, Islam M, et al. Inactivation of Listeria and E. coli by Deep-UV LED: effect of substrate conditions on inactivation kinetics. Sci Rep. 2020;10(1):3411. https://doi.org/10.1038/s41598-020-60459-8 PMID: 32099043

6. Inagaki H, Saito A, Sugiyama H, Okabayashi T, Fujimoto Sh. Rapid inactivation of SARS-CoV-2 with deep-UV LED irradiation. Emerg Microbes Infect. 2020;9(1):1744–1747. https://doi.org/10.1080/2222175 1.2020.1796529 PMID: 32673522

7. Coohill Th P, Sagripanti J-L. Overview of the Inactivation by 254 nm Ultraviolet Radiation of Bacteria with Particular Relevance to Biodefense. Photochem Photobiology. 2008;84(5):1084–1090. https://doi.org/10.1111/j.1751-1097.2008.00387.x PMID: 18627518

8. Hoyer O. Testing performance and monitoring of UV systems for drinking water disinfection. Water Supply. 1998;16(1/2):419–442. 9. Kowalski W. Ultraviolet Germicidal Irradiation Handbook. Berlin, Heidelberg: Springer; 2009. https://doi.org/10.1007/978-3-642-01999-9

9. Zemke V, Podgorsek L, Schoenen D. Ultraviolet disinfection of drinking water. 1.Communication: Inactivation of E. coli and coliform bacteria. Zentralbl Hyg Umweltmed. 1990;190(1/2):51–61. PMID: 2203373

10. Giese N, Darby J. Sensitivity of microorganisms to different wavelengths of UV light: Implications on modeling of medium pressure UV systems. Water Research. 2000;34(16):4007–4013. https://doi.org/10.1016/S00431354(00)00172-X

11. Masjoudi M, Mohseni M, Bolton J R. Sensitivity of Bacteria, Protozoa, Viruses, and Other Microorganisms to Ultraviolet Radiation. J Research of NIST. 2021;126:126021. https://doi.org/10.6028/jres.126.021

12. Chevrefils G, Caron É, Wright H, Sakamoto G, Payment P, Barbeau B, et al. UV Dose Required to Achieve Incremental Log Inactivation of Bacteria, Protozoa and Viruses. IUVA News. 2006;8(1):38–44.

13. Ultraviolet Light Disinfection Data Sheet. ClorDiSys, Rev.122020. Available at: https://www.clordisys.com/pdfs/misc/UV%20Data%20Sheet.pdf [Accessed 31 Aug, 2023]


Review

For citations:


Kamrukov A.S., Chernenkaya T.V., Volodin L.Yu., Petrikov S.S., Popugaev K.A., Bagrov V.V., Bukhtiyarov I.V., Zibarev E.V., Semenov K.A., Krylov V.I. Germicidal Efficiency of 272 nm LED in Relation to the Hospital Strain of Klebsiella Pneumoniae. Russian Sklifosovsky Journal "Emergency Medical Care". 2023;12(3):376-385. https://doi.org/10.23934/2223-9022-2023-12-3-376-385

Views: 543


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 2223-9022 (Print)